Class II Type B2 Fully Automatic Model

BIO SAFETY CABINET CLASS II TYPE B2
FULLY AUTOMATIC MICROPROCESSOR CONTROLLED
LCD/LED display of Air Velocity, Alarm for Undershoot / Overshoot of Air Velocity.
(EN 12469 CERTIFIED EUROPEAN STANDARD)

Principle

Waiometra Bio Safety Cabinet are based on the principle to provide a work area that is completely bathed in a high efficiency perfect air, which is free from any kinds of particulate contamination. These biosafety cabinets are indispensable when the user is working with potentially infections bio hazardous cabinets are indispensable when the user is working with potentially infections bio hazardous agents, bacteria and viruses WAIOMETRA Bio Safety Cabinets ensure the triple protection to the personnel, environment and the product by trapping the potentially harmful micro organisms inside the filters & virus burn out unit.

Class II Type B2 Fully Automatic Model
Display screen to provide information about Air exhaust value, Laminar flow value, real time counter for exhaust & laminar fan, UV elapse hour counter and Filter lifespan Safety alarm for low exhaust, low laminar flow, fault of exhaust fan, fault of laminar flow fan, closing condition of visible window..
Programmable UV Light to allow specific exposure time from 0 to 60 minute with automatic shut-off features UV interlocking with front window i.e. closed automatically when front window is open.
Fully Automatic Microprocessor Controlled.
Automatically adjustment of motor speed to ensure safe airflow.
LCD/LED display of Air Velocity, Alarm for Undershoot / Overshoot of Air Velocity
Password protection (Programmable) for authorised operation.
Initial self cleaning cycle with programmable time.
UV timer showing used time and balance available life in hours
Total HEPA Filter timer showing used time and balance available life in hour.
HEPA filter alarm after HEPA filter life is over and guideline for further action through LCD controller.
Audio Visual door unsafe alarm, if door is at incorrect operating height.
UV light/Door/Fluorescent Light Inter Lock i.e. UV switches ON only when door is closed, and automatically switches OFF fluorescent lights.
If UV light is ON & the operator tries to open the door, it raises audio visual alarm.
Display of safe / unsafe position for door.
Ergonomic Design.
Versatile Usage.
Energy Efficient Mechanism.
Low Noise and Vibration Levels.
Long Life and Low maintenance .
Guaranteed supply of spares for 10 years.
Conforms to ISO 14644.1 Class 3 or better cleanliness level.
Onsite installation and after sales service by company trained engineers.

Waiometra Bio Safety Cabinet is suitable for handling medium to low risk organisms encountered in research investigation, Labs in all branches of medical, Agriculture, Pharmaceuticals & Sciences.

Airflow Pattern : No re-circulation 100% air Exhausted through HEPA/ULPA & Ducted
Model
220-240 VAC, 50/60Hz
FAMB-3 2B FAMB-4 2B FAMB-6 2B
FSSB-3 2B FSSB-4 2B FSSB-6 2B
External Dimensions (WxDxH) 975x875x2175 mm
(38.3"x34.4"x85.6")
1275x875x2175 mm
(50.1"x34.4"x85.6")
1950x875x2175mm
(78"x34.4"x85.6")
Internal Dimensions (WxDxH) 900x600x600mm
(3' x 2' x 2')
1200x600x600mm
(4' x 2' x 2')
1800x600x600mm
(6' x 2' x 2')
Usable Work Area 0.42 m2 (4.5 ft2) 0.56 m2(6.0 ft2) 0.70 m2(7.5 ft2)
Tested Opening 203 mm (8")
Inflow Air Velocity 0.53 m/s (105 fpm)
Downlow Air Velocity 0.30 m/s (60 fpm)
HEPA/ULPA Filter Typical
Efficiency
>99.999% at 0.1 to 0.3 micron, ULPA
>99.999% at MPPS, H14 as per EN 1822 EU
>99.99% on 0.3 micron size particles by DOP test
(filter leakage test according to EN12469>
Air Cleanliness ISO 14644.1 Class 3 or better
Sound Emission per EN 12469 < 60 dBA and vibration free
Fluorescent Lamp Intensity ≥ 1200 lux
Power Supply 220-240 VAC, 50/60Hz
Cabinet Construction Main Body Made of 18 gauge electro-galvanized steel, with oven backed epoxy powder coated finish
Stainless Steel 304 grade
Work Surface Made of single piece Stainless steel 304 grade
Side Wall Made of single piece Stainless Steel 304 grade
Armrest Comfortable armrest Made of Stainless Steel 304 grade for
elevating the operators arms to prevent inflow grille
blockage for safety work
Front Window Frameless front window with non- reflecting safety glass. 
Maximum height of front widow opening is 500mm with 10°
slanted to offer operator comfort while working for long time.
5mm UV resistant with adjustable balancer suspension
Chamber Dynamic chamber plenum design for quiet, uniform airflow,
negative pressure plenum surrounded contaminated positive
pressure plenum
Motor ECM (electro-compensated) Motor automatically adjust
the airflow speed without the use of damper to ensure
continuous safe working conditions
Support stand Provided with leveling screws adjustable upto 100mm approx.
Standard Accessories

Anti-explosion, anti-splash power socket, water/media and gas inlet, Pressure manometer, Gas inlet nozzle, floor leveling screws, Support Stand, castor wheels fixed foot & foot master, UV lamp, electrical outlet sockets and movable stand with wheel & breaks (antimicrobial coated), wire cord & spillage tray with outlet.

(a) Virus Burn out at exhaust (Pls. suffix VB for Cat No.)
(b) PVC duct of 150mm dia per running ft. (Pls. suffix ED for Cat No.)
(c) Magnahelic Gauge (Pls. suffix MG for Cat No.)
(d) Motor Blower to be switched off as soon as the door is fully closed (Please Suffix 0 for cat no.)
(e) Lab Chair/Saddle Stool
(f) Foot Rest.
(g) Gas Burner.
(h) Two or more Sockets of 5 / 15 AMP. Inside chamber.
(i) UPS with battery backup.
(j) IQ/OQ/PQ/DQ certificates
(k) Onsite Particle Counting Test with Air velocity & other tests
EN 12469 Compliant Certificate
ISO - 14644-1
ISO 9001:2015
ISO 13485:2016 Certificate
IEC 61010 -1-2:2001/IEC-60601
CE Certificate
WHO GMP
IQ/OQ/PQ/DQ Certificates (Optional)
Factory Testing Certificate for
(a) Particle Counting which prove the Class 100 working of the instruments (Optional)
(b) Air velocity test certificate (Optional)
(c) Lighting intensity certificate.
(d) Working Manual.
(e) HEPA filter leak test.
(f) Down flow velocity test.
(g) Inflow velocity test.
(h) Airflow smoke pattern test.